Specific Content for Unchained Labs
Acetone Liquid Filter for testing the Stray Light, (measured against air)
Certification Wavelength: 260, 280, 300nm, Cut Off 325nm Slit Width 2 nm
Potassium Dichromate Liquid Filters, (measured against Perchloric Acid Reference Filter)
20 mg/L (0.1-0.3 Abs), 60 mg/L (0.3-0.9 Abs), 120mg/L (0.6-1.8 Abs), HClO4 Reference filter
235, 257, 313, 350nm, Slit Width 2nm
Toluene in hexane Liquid Filter, (measured against hexane Reference Filter)
Scan from 265 to 270nm, Slit Width 1.0, 1.5, 2.0, 2.5, 3.0 nm
Holmium Oxide Liquid Filter for testing the wavelength accuracy (measured against Perchloric Acid Reference Filter)
241, 287, 361, 452, 485, 536, 640nm, Slit Width 2 nm
Recertification Interval Recommended every 18 months
Made to order, available in 3-6 weeks
Product Specifications